2014 Biomedizinische Technik. Biom…

Tightening of healing abutments: influence of torque on bacterial proliferation risk, an in vitro investigation.

, , , ,

Biomedizinische Technik. Biomedical engineering Vol. 59 (6) : 495-500 • Dec 2014

BACKGROUND: Gap at the implant-healing abutment junction can increase the risk of bacterial proliferation. In this study, we determined the leakage at the microgap, and we evaluated hand screwing among clinicians. MATERIALS AND METHODS: The torques tested with nitrogen gas flow were 10, 15, 20, and 30 N cm, and 54 clinicians were asked to torque down a healing abutment as for a surgical procedure. RESULTS: There were no significant differences between 10 and 15 N cm, with a total lack of tightness. For 20 and 30 N cm, there was a notable decrease in leakage. The torque achieved by hand was <10 N cm for 61.7% of the clinicians, between 10 and 15 N cm for 29.1%, between 15 and 20 N cm for 8.0%, and between 20 and 25 N cm for 1.2%. CONCLUSION: There was a significant difference related to the strength of tightening. Under the conditions of our experiment, the gap of connection was reduced with a torque of >/=20 N cm. Only a small portion of the clinicians could obtain these values by hand. Therefore, a dynamometrical manual wrench should be used to minimize the gap during the osseointegration period.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
+2 more
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.