2015 The International journal of …

Reliability and failure modes of two Y-TZP abutment designs.

, , ,

The International journal of prosthodontics Vol. 28 (1) : 75-8 • Jan 2015

Fracture strength and accelerated fatigue reliability of two zirconia abutment systems were tested. Thirty-six implants with a Morse taper (MT; n = 18) or cone (C; n = 18) design were restored with metallic crowns. Loads were applied as single load to failure (SLF) or mouth-motion cycles using a step-stress accelerated life testing (SALT) method. SLF mean values were 690 +/- 430 N and 209 +/- 25 for MT and C groups, respectively. In terms of the SALT results, 8 specimens survived (50,000 cycles) and 7 failed (maximum load 400 N) in the MT group; whereas for the C group all abutments failed before the maximum number of cycles. Failure mode was fracture of the Y-TZP abutments for both groups. Higher reliability for a mission of 50,000 cycles at 175 N for MT versus C designs was determined, and significant differences in fracture modes were observed.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.