[Implantable Cardioverter Defibrillator and Perioperative Magnet Application: A Case Report].
An implantable cardioverter defibrillator (ICD) can falsely recognize noise by monopolar electrocautery as tachyarrhythmia and deliver inappropriate antitachycardia therapy. Application of a clinical magnet on an ICD suspends antitachycardia therapy, but it has not been widely used for this purpose. A 67-year-old male underwent laryngopharyngectomy, cervical esophagectomy, right neck dissection, tracheostomy and reconstruction with free jejunal transplant for recurrent hypopharyngeal cancer. He had an ICD (PARADYM DR8550, Sorin) implanted below the left clavicle for ventricular tachycardia and prolonged QT syndrome. During the operation, a clinical magnet was left on the ICD to disable antitachycardia therapy. The magnet mode of the ICD provided asynchronous AAI pacing at 96 beats x min(-1). The surgery proceeded uneventfully. No episode of ventricular tachyarrythmia or pacing inhibition by electromagnetic interference was observed on electrocardiogram. This case illustrated the potential role of a clinical magnet as an alternative to reprogramming of an ICD by a programmer in the perioperative management of a patient with an ICD when a technical expert to operate a programmer is not available.
No clinical trial protocols linked to this paper
Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.PICO Elements