Load to failure of different zirconia implant abutments with titanium components.

, , , ,

The Journal of prosthetic dentistry Vol. 117 (6) : 749-754 • Jun 2017

STATEMENT OF PROBLEM: Abutments with a zirconia superstructure and a titanium insert have recently become popular. Although they have been tested under static load, their performance under simulated mastication is not well known. PURPOSE: The purpose of this in vitro study was to compare the cyclic load to failure of 3 types of zirconia abutments with different mechanisms of retention of the zirconia to the titanium interface. MATERIAL AND METHODS: Fifteen implants (n=5 per system) and abutments (3 groups: 5 friction fit [Frft]; 5 bonded; and 5 titanium ring friction fit [Ringfrft]) were used. Abutments were thermocycled in water between 5 degrees C and 55 degrees C for 15000 cycles and then cyclically loaded for 20000 cycles or until failure at a frequency of 2 Hz by using a sequentially increased loading protocol up to a maximum of 720 N. The load to failure for each group was recorded, and 1-way analysis of variance was performed. RESULTS: The mean load-to-failure values for the Frft group was 526 N, for the Bond group 605 N, and for the Ringfrft group 288 N. A statistically significant difference was found among all abutments tested (P<.05). CONCLUSIONS: Abutments with the bonded connection showed the highest load-to-failure value, and the abutment with the titanium ring friction fit connection showed the lowest load-to-failure value.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.