2017 Dento maxillo facial radiology

Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV.

, , ,

Dento maxillo facial radiology Vol. 46 (3) : 20160366 • Mar 2017

OBJECTIVES: To evaluate the action of a metal artefact reduction (MAR) tool when artefact-generator metal object is at different positions in the field of view (FOV). METHODS: A cylindrical utility wax phantom, with a metal alloy sample inside, was made. The phantom was positioned centrally and peripherally in the FOV for image acquisition, with and without the MAR tool activation. The standard deviation values (image noise levels) from areas around the metal sample and the control area were obtained. The numbers were compared by Student's t-test (alpha = 0.05). RESULTS: When the tool was activated, a significant difference of image noise was observed for central and peripheral positioning, for both control area (p = 0.0012) and metal area (p = 0.03), and a smaller level of noise was observed for images with phantoms in central positioning. A decrease in image noise with the tool activated was found only in phantoms with the metal object positioned centrally in the FOV. CONCLUSIONS: For the MAR tool to be effective, the artefact-generator object needs to be in the central region of the FOV.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.