2018 American journal of orthodont…

Predictable method to deliver physiologic force for extrusion of palatally impacted maxillary canines.

, , ,

American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics Vol. 153 (2) : 195-203 • Feb 2018

INTRODUCTION: Orthodontic treatment of palatally impacted maxillary canines raises many difficulties; to minimize complications, careful planning of orthodontic extrusion and the use of physiologic force are crucial. The aim of this study was to quantitatively evaluate a simple and reproducible system for orthodontic extrusion of impacted canines that can provide the correct amount of force. METHODS: Ten specimens were constructed, consisting of a cantilever made with a 0.6-mm or 0.7-mm stainless steel wire modeled around a transpalatal bar with 3, 5, or 7 loops in the shape of a helical torsion spring. A mechanical testing machine was used to measure the force produced by the cantilever at 3, 6, 9, 12, and 15 mm of activation. RESULTS: The force values ranged from 1.24 +/- 0.13 N for the 0.7-mm wire with 3 loops to 0.48 +/- 0.04 N for the 0.6-mm wire with 7 loops. The forces measured for the 0.6-mm wire with 3 loops and the 0.7-mm wire with 7 loops were similar at 15 mm of deflection. CONCLUSIONS: The proposed system has a simple and robust design, is easy to construct and manage, and can provide the desired amount of force by changing the wire diameter and number of loops.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.