In vitro comparison of resonance frequency analysis devices to evaluate implant stability of narrow diameter implants at varying drilling speeds in dense artificial bone blocks.

, , , , ,

Clinical implant dentistry and related research Vol. 21 (5) : 1023-1027 • Oct 2019

BACKGROUND: There are no studies that have assessed the implant stability quotient (ISQ) values of narrow diameter implants placed in artificial dense bone blocks at varying drilling speeds (DSs). PURPOSE: The aim of the present in vitro experiment was to compare the performance of OSSTELL and Penguin devices to evaluate implant stability at DSs of 800 and 2000 rpm. MATERIALS AND METHODS: A total of 360 osteotomies were created in dense artificial bone blocks at DSs of 800 and 2000 rpm. Dental implants from three manufacturers (group-1: NobelActive implants, Nobel Biocare, Yorba Linda, California; group-2: Zimmer, Eztetic-Zimmer implants, Zimmer Biomet Dental, Palm Beach Gardens, Florida; and group-3: Astra Tech implant system, Dentsply Sirona, York, Pennsylvania) were randomly placed in these osteotomies using an insertion torque of 15 Ncm (60 implants/group). Implant stability in all bone blocks immediately following implant placement was evaluated using the OSSTELL and Penguin devices. ISQ values were presented as means +/- SD. Statistical significance was set at P < .05. RESULTS: There was no significant difference in the ISQ values obtained from the OSSTELL and Penguin devices for implants in groups 1, 2, and 3. There was no significant difference when ISQ values obtained from the OSSTELL device were compared with the Penguin device for narrow diameter dental implants placed in dense bone blocks with osteotomies performed at 800 and 2000 rpm. ISQ values showed statistically significant higher values for OSSTELL compared to Penguin device. CONCLUSION: The OSSTELL and Penguin devices are reliable for the assessment of implant stability in dense artificial bone. Implant design and site-DS does not seem to have a significant impact of implant stability in artificial dense bone blocks.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.