2020 Journal of endodontics

Mechanical Performance and Metallurgical Features of ProTaper Universal and 6 Replicalike Systems.

, , , , , , ,

Journal of endodontics Vol. 46 (12) : 1884-1893 • Dec 2020

INTRODUCTION: This study aimed to compare the ProTaper Universal (PTU; Dentsply Maillefer, Ballaigues, Switzerland) system with 6 replicalike instruments regarding instrument design, mechanical performance, and alloy characteristics. METHODS: New rotary instruments (size 20/.07v) of PTU and 6 replicalike systems (EdgeTaper [EdgeEndo, Johnson City, TN], U-File [Dentmark, Ludhiana, India], Go-Taper Universal [Access, Shenzhen, China], Super Files [Flydent, Shenzhen, China], Multitaper [Proclinic Expert, Besancon, France], and Pluri Taper [Bestdent, Shenzhen, China]) (n = 329) were selected and evaluated regarding their design, mechanical performance, and metallurgical characteristics. The results were compared using the nonparametric Mood median test and 1-way analysis of variance with the significance level set at 0.05. RESULTS: Replicalike instruments were similar to PTU regarding design, maximum torque to fracture (P > .05), and ratio of nickel and titanium elements. Scanning electron microscopy revealed differences in the geometry of the tip and finishing surfaces. Time to fracture of Go-Taper Universal (50.0 seconds) was similar to PTU (44.0 seconds) (P > .05), but the U-File (63.5 seconds), Edge Taper (87.5 seconds), and Super Files (130 seconds) showed significantly higher values (P < .05). The highest maximum torque was observed in the Super Files instrument (P < .05), but no statistical difference was noted among the other instruments (P > .05). The angle of rotation before rupture of the Multitaper (574 degrees ) and Pluri Taper (481 degrees ) was higher than PTU (354 degrees ) (P < .05). Differential scanning calorimetric analysis showed clear differences on the transformation temperatures among the tested instruments. The R-phase start and finish temperatures of PTU were 11.0 degrees C and -18.0 degrees C, respectively. CONCLUSIONS: Overall, replicalike instruments were similar to PTU regarding the torsional resistance, geometric design, and near equiatomic proportions of nickel and titanium elements. In 3 of the 4 mechanical tests, the EdgeTaper (torque, angle of rotation, and bending load) and Go-Taper Universal (time to fracture, torque, and angle of rotation) systems were similar to PTU.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.