2022 The British journal of oral &…

In vitro comparison of fixation methods used in sagittal split osteotomy with a major advancement and counterclockwise rotation.

,

The British journal of oral & maxillofacial surgery Vol. 60 (5) : 617-622 • Jun 2022

The aim of this study was to compare the mechanical characteristics of five commonly used fixation methods for sagittal split osteotomy with a major advancement and counterclockwise rotation with a model that mimics the human mandible. Twenty-five virtually osteotomised and 3-dimensionally printed hemimandibles were fixed with an 8mm gap at the upper border and a 12mm gap at the lower border of the osteotomy. The following fixation methods were used: Group 1: a single miniplate, group 2: two miniplates, group 3: three bicortical screws in the upper border, group 4: three bicortical screws in an inverted L configuration, group 5: a single miniplate and a bicortical screw. A mechanical test using vertical linear loading at 1 mm/min velocity on a universal testing machine was carried out. The forces needed to displace the distal segment from 1-5 mm were recorded. The difference between groups was statistically significant for all displacement levels. Group 1 showed the minimum, group 5 showed the maximum resistance in almost all displacement values and there was a trend for higher to lower resistance values towards groups 5 to 1. Fixation methods that used one bicortical screw and a miniplate or three bicortical screws showed higher load resistance against displacement.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.