2023 Journal of cranio-maxillo-fac…

Mandibular resection and defect reconstruction guided by a contour registration-based augmented reality system: A preclinical trial.

, , , , , , , , ,

Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery Vol. 51 (6) : 360-368 • Jun 2023

The aim of this study was to verify the feasibility and accuracy of a contour registration-based augmented reality (AR) system in jaw surgery. An AR system was developed to display the interaction between virtual planning and images of the surgical site in real time. Several trials were performed with the guidance of the AR system and the surgical guide. The postoperative cone beam CT (CBCT) data were matched with the preoperatively planned data to evaluate the accuracy of the system by comparing the deviations in distance and angle. All procedures were performed successfully. In nine model trials, distance and angular deviations for the mandible, reconstructed fibula, and fixation screws were 1.62 +/- 0.38 mm, 1.86 +/- 0.43 mm, 1.67 +/- 0.70 mm, and 3.68 +/- 0.71 degrees , 5.48 +/- 2.06 degrees , 7.50 +/- 1.39 degrees , respectively. In twelve animal trials, results of the AR system were compared with the surgical guide. Distance deviations for the bilateral condylar outer poles were 0.93 +/- 0.63 mm and 0.81 +/- 0.30 mm, respectively (p = 0.68). Distance deviations for the bilateral mandibular posterior angles were 2.01 +/- 2.49 mm and 2.89 +/- 1.83 mm, respectively (p = 0.50). Distance and angular deviations for the mandible were 1.41 +/- 0.61 mm, 1.21 +/- 0.18 mm (p = 0.45), and 6.81 +/- 2.21 degrees , 6.11 +/- 2.93 degrees (p = 0.65), respectively. Distance and angular deviations for the reconstructed tibiofibular bones were 0.88 +/- 0.22 mm, 0.84 +/- 0.18 mm (p = 0.70), and 6.47 +/- 3.03 degrees , 6.90 +/- 4.01 degrees (p = 0.84), respectively. This study proposed a contour registration-based AR system to assist surgeons in intuitively observing the surgical plan intraoperatively. The trial results indicated that this system had similar accuracy to the surgical guide.

No clinical trial protocols linked to this paper

Clinical trials are automatically linked when NCT numbers are found in the paper's title or abstract.
PICO Elements

No PICO elements extracted yet. Click "Extract PICO" to analyze this paper.

Paper Details
MeSH Terms
Associated Data

No associated datasets or code repositories found for this paper.

Related Papers

Related paper suggestions will be available in future updates.