2023 The International journal of …

Comparative Mechanical Testing for Digitally Produced Provisional Fixed Partial Dentures vs the Conventional Method: An In Vitro Study.

, , , , ,

The International journal of prosthodontics Vol. 36 (6) : 748-753 • Dec 2023

PURPOSE: To examine and compare the fracture strength of implant-cemented fixed partial denture (FPD) prostheses fabricated with digital vs conventional chairside methods. MATERIALS AND METHODS: Three groups of seven specimens each were produced: group A (3D printing); group B (milling); and group C (conventional chairside manufacturing), which served as a control. All groups were cemented to standard implant abutments placed in artificial bone blocks. Fracture strength testing was performed using a universal testing machine. Statistical analysis of the resultant maximum forces was performed using SPSS version 25 software (Mann- Whitney U test, P < .05). RESULTS: The mean fracture load value of the group A FPDs was 260.14 N +/- 28.88, for group B was 663.57 N +/- 140.55, and for group C was 266.65 N +/- 63.66. CONCLUSIONS: Milled provisional FPDs showed a higher fracture resistance compared to 3D-printed and control groups. However, no such difference could be detected between the 3D-printed and control groups.